Apertura
|
Measurement aperture: 8 mm (0.31?) diameter (effective measurement aperture during scanning is depending on the patch size and measurement speed)
|
Espacio en disco disponible
|
2GB
|
Calibración
|
Manual on external ceramic white reference (included)
|
Interfaz de comunicación
|
USB 1.1
|
Conectividad
|
Powered USB port
|
Dimensiones (largo, ancho, alto)
|
i1Pro device: 162 mm x 69 mm x 64 mm (6.4? x 2.7? x 2.5?)
|
Resolución de la pantalla
|
1024x768 pixels or higher
|
Nivel de experiencia
|
Intermediate to Advanced
|
Humedad
|
30 to 85% RH (non-condensing)
|
Tamaño del área de iluminación
|
12 mm (0.47?)
|
Acuerdo entre instrumentos
|
0.4 ?E00* average, 1.0 ?E00* max. (deviation from X-Rite manufacturing standard at a temperature of 23ºC (73.4ºF) on 12 BCRA tiles (D50, 2º))
|
Conexión a Internet
|
Required for software install, download and automatic update
|
Idiomas compatibles
|
Chinese (Simplified), English, French, German, Italian, Japanese, Korean, Portuguese, Spanish
|
Fuente de luz
|
i1Pro3 Plus LED Illumination (including UV)
|
Macintosh
|
MacOS X 10.11x, 10.12x or 10.13x (with latest upgrades installed)
|
Condiciones de medición
|
? M0 - UV included - ISO 13655:2017
? M1 - D50 - ISO 13655:2017
? M2 - UV excluded Filter - ISO 13655:2017
? M3 - Polarization - ISO 13655:2017
|
Frecuencia de la medición en el modo de escaneo
|
400 measurements per second
|
Geometría de medición
|
45°:0° ring illumination optics, ISO 13655:2017
|
Zona de medición
|
8mm (0.315?)
|
Media Thickness
|
1mm or less
|
Memoria
|
2GB Minimum
|
Cantidad de pantallas admitidas por estación de trabajo
|
Up to 4
|
Rango de temperatura de operación
|
10°C to 35°C
50°F to 95°F
|
Tamaño del parche
|
Minimal Patch Size in Scanning Mode: 16 x 16 mm (0.63? x 0.63?) (Width x Height) with sensor ruler 20 x 16 mm (0.79? x 0.63?) (Width x Height) without sensor ruler
|
Rango fotométrico
|
10nm; Sampling interval 3.5nm
|
Procesador
|
MAC: Intel® Core 2 Duo CPU or better; WINDOWS: Intel® Core 2 Duo or AMD Athlon? XP or better
|
Longitud de escaneo
|
Maximum 515mm (20.28")
|
Compatible con el objetivo del escáner
|
Yes
|
Capacidad de escaneo
|
Yes
|
Seguridad
|
User must have Administrator rights to install and uninstall the application
|
Repetitividad de corto plazo: blanco
|
Reflectance: 0.1 ?E94* (D50,2°, mean of 10 measurements every 3 seconds on white)
|
Kit de desarrollo de software
|
Available for OEM customers
|
Analizador de espectro
|
i1® technology with built-in wavelengths check; Holographic diffraction grating with 128-pixel diode array
|
Intervalo espectral
|
380 - 730 nm
|
Presentación del espectro
|
380 ... 730 nm in 10 nm steps
|
Rango de temperatura de almacenamiento
|
-20° to 50°C
-4° to 122°F
|
Formato de perfil compatible
|
ICC version 2, version 4.3, & iccMax
|
Unité de couleur
|
Black/Silver
|
Experiencia de usuario
|
Choose between a ?basic?, wizard-driven interface or an ?advanced?, user-driven interface for highest level professional results
|
Tarjeta de video
|